Two-photon laser scanning microscopy with electrowetting-based prism scanning

Biomed Opt Express. 2017 Nov 7;8(12):5412-5426. doi: 10.1364/BOE.8.005412. eCollection 2017 Dec 1.

Abstract

Laser scanners are an integral part of high resolution biomedical imaging systems such as confocal or 2-photon excitation (2PE) microscopes. In this work, we demonstrate the utility of electrowetting on dielectric (EWOD) prisms as a lateral laser-scanning element integrated in a conventional 2PE microscope. To the best of our knowledge, this is the first such demonstration for EWOD prisms. EWOD devices provide a transmissive, low power consuming, and compact alternative to conventional adaptive optics, and hence this technology has tremendous potential. We demonstrate 2PE microscope imaging of cultured mouse hippocampal neurons with a FOV of 130 × 130 μm2 using EWOD prism scanning. In addition, we show simulations of the optical system with the EWOD prism, to evaluate the effect of propagating a Gaussian beam through the EWOD prism on the imaging quality. Based on the simulation results a beam size of 0.91 mm full width half max was chosen to conduct the imaging experiments, resulting in a numerical aperture of 0.17 of the imaging system.

Keywords: (010.1080) Active or adaptive optics; (170.0110) Imaging systems; (170.2520) Fluorescence microscopy; (180.5810) Scanning microscopy; (230.2090) Electro-optical devices; (230.5480) Prisms.