Estimation and comparison of the radiation effective dose during coronary computed tomography angiography examinations on single-source 64-MDCT and dual-source 128-MDCT

J Radiol Prot. 2017 Dec;37(4):826-836. doi: 10.1088/1361-6498/aa823f.

Abstract

Goal: To estimate and compare the radiation dose associated with coronary computed tomography angiography (CCTA) examinations on two multi-detector CT scanners (MDCT), 64-MDCT and 128-MDCT, in daily practice.

Methods: Scan parameters of 90 patients undergoing retrospective electrocardiographic gating spiral CCTA exam were recorded during a period on a single-source 64-MDCT and a dual-source 128-MDCT, and average scan parameters were derived that were used for dosimetry. The computed tomography dose index (CTDI) with a pencil ionisation chamber and polymethyl methacrylate body phantom with diameter of 32 cm was measured on both scanners. The dose-length product (DLP) was calculated and the DLP to effective dose conversion factor (for chest scan at 120 kV of 0.014 mSv mGy-1 cm-1) was used to estimate effective dose (ED).

Results: Patients' heart rate, scan length, pitch factor, CTDIv, DLP and ED for 128-MDCT were 64 (5) (beats min-1), 161 (10) (mm), 0.26, 47 (12) (mGy), 769 (212) (mGy cm) and 10.3 (3.1) (mSv), respectively [mean (one standard deviation)]. Patients' heart rate, scan length, pitch factor, CTDIv, DLP and ED for 64-MDCT were 60 (7) (beats min-1), 172 (14) (mm), 0.2, 60 (6) (mGy), 1068 (98) (mGy cm) and 14.9 (1.4) (mSv), respectively.

Conclusion: Our results indicated that the CTDIv, DLP and the effective dose with 128-MDCT is significantly lower than with 64-MDCT (p < 0.05). As differences between the exposure parameter mAs on two CT scanners was not significant (p > 0.05) and the kV was constant for both scanners (120 kV), the differences resulted from a shorter scan length on the 128-MDCT and use of a higher pitch factor (0.26 and 0.2 in the 128-MDCT and 64-MDCT, respectively). Comparison with other published studies confirms the findings and indicates methods for reducing patient dose.

MeSH terms

  • Cardiac-Gated Imaging Techniques
  • Computed Tomography Angiography / instrumentation*
  • Coronary Angiography*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Models, Anatomic
  • Phantoms, Imaging
  • Radiation Dosage*
  • Radiometry / methods*