    J Chromatogr. 1989 Mar 31;465(2):285-96.

    Trace detection of modified DNA bases via moving-belt liquid chromatography-mass spectrometry using electrophoric derivatization and negative chemical ionization.

    Annan RS, Kresbach GM, Giese RW, Vouros P.

    Department of Chemistry and Barnett Institute, Northeastern University, Boston, MA 02115.

    Electrophoric derivatives of 5-methylcytosine and 5-hydroxymethyluracil nucleobases are determined using high-performance liquid chromatography-mass spectrometry coupled via a moving-belt interface. Standards as well as samples derived from DNA are analysed. As little as 9.9 pg (signal-to-noise ratio 5) and 180 fg (signal-to-noise ratio 10) of the respective nucleobases are detected in the electron-capture negative chemical ionization mode, and linear responses are observed over a moderate dynamic range. In a comparison study, liquid chromatography-electron-capture negative chemical ionization mass spectrometry is found to have a sensitivity comparable to gas chromatography-electron-capture negative chemical ionization mass spectrometry for 5-hydroxymethyluracil. A detection limit of 60 fg (signal-to-noise ratio 5) by gas chromatography-mass spectrometry is only three-fold better than the amount detected by liquid chromatography-mass spectrometry using the same mass spectrometer.

    PMID: 2745598 [PubMed - indexed for MEDLINE]