Predicting Out-of-Office Blood Pressure in the Clinic (PROOF-BP): Derivation and Validation of a Tool to Improve the Accuracy of Blood Pressure Measurement in Clinical Practice

Hypertension. 2016 May;67(5):941-50. doi: 10.1161/HYPERTENSIONAHA.115.07108. Epub 2016 Mar 21.

Abstract

Patients often have lower (white coat effect) or higher (masked effect) ambulatory/home blood pressure readings compared with clinic measurements, resulting in misdiagnosis of hypertension. The present study assessed whether blood pressure and patient characteristics from a single clinic visit can accurately predict the difference between ambulatory/home and clinic blood pressure readings (the home-clinic difference). A linear regression model predicting the home-clinic blood pressure difference was derived in 2 data sets measuring automated clinic and ambulatory/home blood pressure (n=991) using candidate predictors identified from a literature review. The model was validated in 4 further data sets (n=1172) using area under the receiver operator characteristic curve analysis. A masked effect was associated with male sex, a positive clinic blood pressure change (difference between consecutive measurements during a single visit), and a diagnosis of hypertension. Increasing age, clinic blood pressure level, and pulse pressure were associated with a white coat effect. The model showed good calibration across data sets (Pearson correlation, 0.48-0.80) and performed well-predicting ambulatory hypertension (area under the receiver operator characteristic curve, 0.75; 95% confidence interval, 0.72-0.79 [systolic]; 0.87; 0.85-0.89 [diastolic]). Used as a triaging tool for ambulatory monitoring, the model improved classification of a patient's blood pressure status compared with other guideline recommended approaches (93% [92% to 95%] classified correctly; United States, 73% [70% to 75%]; Canada, 74% [71% to 77%]; United Kingdom, 78% [76% to 81%]). This study demonstrates that patient characteristics from a single clinic visit can accurately predict a patient's ambulatory blood pressure. Usage of this prediction tool for triaging of ambulatory monitoring could result in more accurate diagnosis of hypertension and hence more appropriate treatment.

Keywords: ambulatory blood pressure monitoring; hypertension; masked hypertension; white coat hypertension.

Publication types

  • Comparative Study
  • Multicenter Study
  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Adult
  • Aged
  • Algorithms
  • Blood Pressure Determination / methods*
  • Blood Pressure Monitoring, Ambulatory / methods*
  • Canada
  • Circadian Rhythm*
  • Cohort Studies
  • Databases, Factual
  • Female
  • Humans
  • Linear Models
  • Male
  • Masked Hypertension / diagnosis*
  • Middle Aged
  • Office Visits
  • Predictive Value of Tests
  • ROC Curve
  • Risk Assessment
  • Sensitivity and Specificity
  • United Kingdom
  • United States
  • White Coat Hypertension / diagnosis*