The analysis was implemented concerning diagnostic parameters of commercial quick tests (immune chromatographic tests BinaxNOW Influenza A&B and BinaxNow RSV Alere, Scarborough Inc., USA) under detection of antigens of influenza virus A and respiratory syncytial virus in clinical materials. The polymerase chain reaction in real-time and isolation ofviruses in cell cultures. The analysis of naso-pharyngeal smears from 116 children demonstrated that sensitivity and specifcity of detection of influenza virus A using device mariPOC in comparison with polymerase chain reaction made up to 93.8% and 99.0% correspondingly at total concordance of results of both techniques as 98.3%. At diagnosing of respiratory syncytial virus using device mariPOC parameters made up to 77.3%, 98.9% and 862% as compared with polymerase chain reaction. The sensitivity, specificity and total concordance of results of immune chromatographic tests BinaxNOW in comparison ofpolymerase chain reaction made up to 86.7%, 100% and 96.2% correspondingly at detection of influenza virus A and 80.9%, 97.4% and 91.6% correspondingly at detection of respiratory syncytial virus. In comparison with isolation technique in cell cultures sensitivity of system mariPOC and immune chromatographic tests proved to be in 1.3-1.4 times higher at detection of influenza virus A and in 1.7-2 times higher in case of isolation of respiratory syncytial virus. There is no statistically significant differences between diagnostic parameters received for mariPOC and immune chromatographic tests at diagnosing influenza virus A and respiratory syncytial viral infection.