Improvement of the Off-Resonance Saturation, an MRI sequence for positive contrast with SPM particles: Theoretical and experimental study

J Magn Reson. 2016 Apr:265:99-107. doi: 10.1016/j.jmr.2016.02.002. Epub 2016 Feb 9.

Abstract

The SuperParaMagnetic particles (SPM particles) are used as contrast agents in MRI and produce negative contrast with conventional T2 or T2(∗)-weighted sequences. Unfortunately, the SPM particle detection on images acquired with such sequences is sometimes difficult because negative contrast can be created by artifacts such as air bubbles or calcification. To overcome this problem, new sequences as Off-Resonance Saturation (ORS) were developed to produce positive contrast with SPM particles. This work explores a new way to optimize the contrast generated by the ORS sequence by increasing the number of saturation pulses applied before the imaging sequence. This modified sequence is studied with numerical simulations and experiments on agarose gel phantoms. A theoretical model able to predict the contrast for different values of the sequence parameters is also developed. The results show that the contrast increases with the saturation pulses number with an optimal value of three saturation pulses in order to avoid artifacts and limit the Specific Absorption Rate (SAR) effect. The dependence of the contrast on the SPM particle concentration and sequence parameters is comparable to what was observed for the ORS sequence.

Keywords: Contrast agents; Iron oxide nanoparticles; Magnetic resonance imaging; Off-resonance imaging; Positive contrast imaging; Superparamagnetic nanoparticles.