Non-invasive quantification of hepatic steatosis in living, related liver donors using dual-echo Dixon imaging and single-voxel proton spectroscopy

Clin Radiol. 2016 Jan;71(1):58-63. doi: 10.1016/j.crad.2015.10.002. Epub 2015 Nov 7.

Abstract

Aim: To evaluate the diagnostic implications of hepatic fat fraction calculated using dual-echo Dixon imaging and (1)H magnetic resonance spectroscopy (MRS) to detect hepatic steatosis in potential liver donors using histopathology as the reference standard.

Materials and methods: One hundred and forty-five potential liver donors were included in the study. Magnetic resonance imaging (MRI) was performed using a 1.5 T system using a three-dimensional dual-echo MRI sequence with automated reconstruction of in-phase (IP), out-of-phase (OP), fat-signal-only, and water-signal-only images. Hepatic fat fraction was calculated by drawing 15 regions of interest on the IP, OP, fat-only, and water-only images. Single-voxel MRS was performed at echo times (TEs) of 30 ms in the right and left lobes of liver. Liver fat fraction was calculated from water and fat peaks. One hundred and forty-five biopsies were prospectively evaluated for steatosis by a pathologist using traditional determination of the cell-count fraction. MRI and pathology values of steatosis were correlated using Pearson's correlation coefficient. The sensitivity and specificity of each of these methods was calculated using histopathology as the reference standard. Reproducibility was assessed in 40 patients who had repeat scanning within 4-40 days. Measurement error was calculated from the coefficient of variation (CoV) with histopathologically proven <5% fat (n=112).

Results: The Bland-Altman limits of agreement with 95% confidence intervals (CI) was -2.9 to 5.3%. The intraclass correlation coefficient (ICC) for interobserver variability and reproducibility was 0.94 (95% CI: 0.91-0.97), 0.92 (95% CI: 0.91-0.97). The CoV was 7.6% (95% CI: 3.4-11.85). The area under the receiver operating characteristic (ROC) curve (AUC) for Dixon imaging 0.89 (95% CI: 0.87-0.91), for MRS 0.88 (95% CI: 0.86-0.90). The sensitivity for detecting <5% fat was 84% and specificity was 90%.

Conclusion: Combination of dual-echo Dixon imaging and proton MRS is a useful tool for the preoperative diagnosis of hepatic steatosis in potential living liver donors. This can help avoid unnecessary biopsies in these patients.