Experimental determination of field factors ([Formula: see text]) for small radiotherapy beams using the daisy chain correction method

Phys Med Biol. 2015 Aug 7;60(15):5813-31. doi: 10.1088/0031-9155/60/15/5813. Epub 2015 Jul 10.

Abstract

Recently, Alfonso et al proposed a new formalism for the dosimetry of small and non-standard fields. The proposed new formalism is strongly based on the calculation of detector-specific beam correction factors by Monte Carlo simulation methods, which accounts for the difference in the response of the detector between the small and the machine specific reference field. The correct calculation of the detector-specific beam correction factors demands an accurate knowledge of the linear accelerator, detector geometry and composition materials. The present work shows that the field factors in water may be determined experimentally using the daisy chain correction method down to a field size of 1 cm × 1 cm for a specific set of detectors. The detectors studied were: three mini-ionization chambers (PTW-31014, PTW-31006, IBA-CC01), three silicon-based diodes (PTW-60018, IBA-SFD and IBA-PFD) and one synthetic diamond detector (PTW-60019). Monte Carlo simulations and experimental measurements were performed for a 6 MV photon beam at 10 cm depth in water with a source-to-axis distance of 100 cm. The results show that the differences between the experimental and Monte Carlo calculated field factors are less than 0.5%-with the exception of the IBA-PFD-for field sizes between 1.5 cm × 1.5 cm and 5 cm × 5 cm. For the 1 cm × 1 cm field size, the differences are within 2%. By using the daisy chain correction method, it is possible to determine measured field factors in water. The results suggest that the daisy chain correction method is not suitable for measurements performed with the IBA-PFD detector. The latter is due to the presence of tungsten powder in the detector encapsulation material. The use of Monte Carlo calculated [Formula: see text] is encouraged for field sizes less than or equal to 1 cm × 1 cm for the dosimeters used in this work.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Computer Simulation
  • Diamond / chemistry
  • Humans
  • Models, Theoretical
  • Monte Carlo Method
  • Particle Accelerators / instrumentation*
  • Phantoms, Imaging*
  • Photons / therapeutic use*
  • Radiometry / instrumentation*
  • Radiometry / methods
  • Radiotherapy / instrumentation*
  • Radiotherapy / methods
  • Silicon / chemistry
  • Water / chemistry*

Substances

  • Water
  • Diamond
  • Silicon