Purity assessment of recombinant human granulocyte colony-stimulating factor in finished drug product by capillary zone electrophoresis

Electrophoresis. 2014 Sep;35(18):2608-14. doi: 10.1002/elps.201300552. Epub 2014 Aug 1.

Abstract

Current methods for determination of impurities with different charge-to-volume ratio are limited especially in terms of sensitivity and precision. The main goal of this research was to establish a quantitative method for determination of impurities with charges differing from that of recombinant human granulocyte colony-stimulating factor (rhG-CSF, filgrastim) with superior precision and sensitivity compared to existing methods. A CZE method has been developed, optimized, and validated for a purity assessment of filgrastim in liquid pharmaceutical formulations. Optimal separation of filgrastim from the related impurities with different charges was achieved on a 50 μm id fused-silica capillary of a total length of 80.5 cm. A BGE that contains 100 mM phosphoric acid adjusted to pH 7.0 with triethanolamine was used. The applied voltage was 20 kV while the temperature was maintained at 25°C. UV detection was set to 200 nm. Method was validated in terms of selectivity/specificity, linearity, precision, LOD, LOQ, stability, and robustness. Linearity was observed in the concentration range of 6-600 μg/mL and the LOQ was determined to be 0.3% relative to the concentration of filgrastim of 0.6 mg/mL. Other validation parameters were also found to be acceptable; thus the method was successfully applied for a quantitative purity assessment of filgrastim in a finished drug product.

Keywords: Capillary zone electrophoresis; Filgrastim; Purity; Recombinant human granulocyte colony-stimulating factor; Validation.

MeSH terms

  • Dosage Forms
  • Drug Contamination
  • Drug Stability
  • Electrophoresis, Capillary / methods*
  • Filgrastim
  • Granulocyte Colony-Stimulating Factor / chemistry*
  • Linear Models
  • Recombinant Proteins / chemistry
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Dosage Forms
  • Recombinant Proteins
  • Granulocyte Colony-Stimulating Factor
  • Filgrastim