Fast shear compounding using robust 2-D shear wave speed calculation and multi-directional filtering

Ultrasound Med Biol. 2014 Jun;40(6):1343-55. doi: 10.1016/j.ultrasmedbio.2013.12.026. Epub 2014 Mar 6.

Abstract

A fast shear compounding method was developed in this study using only one shear wave push-detect cycle, such that the shear wave imaging frame rate is preserved and motion artifacts are minimized. The proposed method is composed of the following steps: 1. Applying a comb-push to produce multiple differently angled shear waves at different spatial locations simultaneously; 2. Decomposing the complex shear wave field into individual shear wave fields with differently oriented shear waves using a multi-directional filter; 3. Using a robust 2-D shear wave speed calculation to reconstruct 2-D shear elasticity maps from each filter direction; and 4. Compounding these 2-D maps from different directions into a final map. An inclusion phantom study showed that the fast shear compounding method could achieve comparable performance to conventional shear compounding without sacrificing the imaging frame rate. A multi-inclusion phantom experiment showed that the fast shear compounding method could provide a full field-of-view, 2-D and compounded shear elasticity map with three types of inclusions clearly resolved and stiffness measurements showing excellent agreement to the nominal values.

Keywords: 2-D shear wave speed; Acoustic radiation force; Comb-push; Directional filter; Shear compounding; Shear wave elastography.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Algorithms
  • Artifacts
  • Elasticity Imaging Techniques / instrumentation
  • Elasticity Imaging Techniques / methods*
  • Image Processing, Computer-Assisted / methods*
  • Phantoms, Imaging
  • Signal-To-Noise Ratio