Radius dependence of FP-CIT quantification: a Monte Carlo-based simulation study.

Author information

  • 1Department of Nuclear Medicine, University of Munich, Marchioninistr. 15, 81377, Munich, Germany, walter.koch@med.uni-muenchen.de.

Abstract

OBJECTIVE:

Dopamine transporter imaging with SPECT is a valuable tool for both clinical routine and research studies. Semi-quantitative analysis plays a key role in interpreting the scans, but is dependent on numerous factors, rotational radius being one of them. This study systematically evaluates the potential influence of radius of rotation on apparent tracer binding and describes methods for correction.

METHODS:

Monte Carlo simulation scans of a digital brain phantom with various disease states and various radii of rotation ranging from 13 to 30 cm were analyzed using 4 different methods of semi-quantification. Different volumes of interest as well as a method with partial volume correction were applied.

RESULTS:

For conventional 3D semi-quantification methods the decrease of measured striatal binding per cm additional radius rotation lied in the range between 2.5 and 3.1 %, whereas effects were negligible when applying recovery-corrected quantification. Effects were independent of disease state.

CONCLUSION:

Partial volume effects with increasing radius of rotation can lead to considerable decrease of measured binding ratios, particularly when applying dopamine transporter imaging in a research setting. Standardization of acquisition radius can avoid the effect; correction seems feasible, but the correction factors depend on the quantification approach applied.

[PubMed - indexed for MEDLINE]