J Med Eng Technol. 2013 Oct;37(7):416-23. doi: 10.3109/03091902.2013.824510. Epub 2013 Aug 27.

Indentation loading response of a resonance sensor--discriminating prostate cancer and normal tissue.

Author information

  • 1Department of Applied Physics and Electronics .

Abstract

Prostate cancer is the most common type of cancer among men worldwide. Mechanical properties of prostate tissue are promising for distinguishing prostate cancer from healthy prostate tissue. The aim was to investigate the indentation loading response of a resonance sensor for discriminating prostate cancer tissue from normal tissue. Indentation measurements were done on prostate tissue specimens ex vivo from 10 patients from radical prostatectomy. The measurement areas were analysed using standard histological methods. The stiffness parameter was linearly dependent on the loading force (average R(2 )= 0.90) and an increased loading force caused a greater stiffness contrast of prostate cancer vs normal tissue. The accuracy of the stiffness contrast was assessed by the ROC curve with the area under the curve being 0.941 for a loading force of 12.8 mN. The results are promising for the development of a resonance sensor instrument for detecting prostate cancer.

[PubMed - indexed for MEDLINE]