Format

Send to:

Choose Destination
See comment in PubMed Commons below
Ann Lab Med. 2013 Jul;33(4):261-7. doi: 10.3343/alm.2013.33.4.261. Epub 2013 Jun 24.

Comparison of the accuracy of noninvasive hemoglobin sensor (NBM-200) and portable hemoglobinometer (HemoCue) with an automated hematology analyzer (LH500) in blood donor screening.

Author information

  • 1Department of Biomedical Laboratory Science, Eulji University College of Health Science, Seongnam, Korea.

Abstract

BACKGROUND:

The Hb levels of prospective blood donors are usually determined using a finger prick test. A new noninvasive Hb device has the advantage of not causing any sampling pain. The purpose of this study was to evaluate the accuracy of the noninvasive Hb sensor and to compare its measurements with those of a currently used portable hemoglobinometer.

METHODS:

Hb was measured using a noninvasive Hb sensor (NBM-200; OrSense, Israel), a portable hemoglobinometer (HemoCue; HemoCue AB, Sweden), and an automated hematology analyzer (LH500; Beckman Coulter, USA). The correlations between Hb measurements taken by the NBM-200 and HemoCue with those by an automated hematology analyzer were assessed using intraclass correlation coefficients (ICCs). Hb measurements were compared among 3 different Hb level groups.

RESULTS:

The mean Hb values of 506 blood donors were 14.1 g/dL by the NBM-200, 14.0 g/dL by the LH500, and 14.3 g/dL by the HemoCue. The correlation between the LH500 and the NBM-200 was substantial (ICC=0.69), while that between the LH500 and the HemoCue agreed almost perfectly (ICC=0.86).

CONCLUSIONS:

The possibility to judge to be eligible for donors who are ineligible to donate was substantial when using NBM-200. Even though the NBM-200 has the apparent advantage of noninvasiveness, its use in pre-screening should be given meticulous attention. Since pre-donation testing is crucial to protecting donors' health, complete evaluation of the instrument should be performed prior to use.

KEYWORDS:

Blood donors; Donor selection; Hemoglobin; Safety management

PMID:
23826562
[PubMed - indexed for MEDLINE]
PMCID:
PMC3698304
Free PMC Article
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for The Korean Society for Laboratory Medicine Icon for PubMed Central
    Loading ...
    Write to the Help Desk