Hemodial Int. 2014 Jan;18(1):172-4. doi: 10.1111/hdi.12054. Epub 2013 May 28.

Comparison of urea clearance in low-efficiency low-flux vs. high-efficiency high-flux dialyzer membrane with reduced blood and dialysate flow: an in vitro analysis.

Author information

  • 1Department of Medicine, University of Washington, Seattle, Washington, USA.

Abstract

Rapid removal of small molecules during hemodialysis places an acutely ill patient with kidney failure at an increased risk of hemodynamic instability and for dialysis disequilibrium syndrome. The use of high-flux, high-efficiency (HEF) dialyzers may increase this risk despite reductions in blood and dialysate flow. We performed in vitro experiments to compare urea clearance at low dialysate flow and various blood flows using a low-efficiency low-flux (LEF) and a HEF membrane. Compared to LEF, there was a significant increase in the clearance of urea at all blood flows with the HEF (all P values < 0.005). HEF dialyzer (F180NR) had higher urea clearance at a blood flow of 150 mL/min than LEF dialyzer (F5) at blood flow of 300 mL/min (144.1 ± 0.99 vs. 130.1 ± 0.001 mL/min for F180 vs. F5, respectively, P < 0.002). Our data suggest that use of HEF dialyzer are not as safe as LEF in high-risk acute dialysis patients since these are associated with more rapid removal of urea despite reduction in blood and dialysate flow as compared to LEF.

© 2013 International Society for Hemodialysis.

KEYWORDS:

Urea; clearance; dialyzer; efficiency

PMID:
23714225
[PubMed - indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Blackwell Publishing
    Loading ...
    Write to the Help Desk