Phys Med Biol. 2013 May 21;58(10):3163-76. doi: 10.1088/0031-9155/58/10/3163. Epub 2013 Apr 19.

Reliability of in vivo measurements of the dielectric properties of anisotropic tissue: a simulative study.

Author information

  • 1School of Biomedical Engineering, The Fourth Military Medical University, Changle West Road No. 169, Xi'an, People's Republic of China.

Abstract

A simulative study was performed to measure the dielectric properties of anisotropic tissue using several in vivo and in vitro probes. COMSOL Multiphysics was selected to carry out the simulation. Five traditional probes and a newly designed probe were used in this study. One of these probes was an in vitro measurement probe and the other five were in vivo. The simulations were performed in terms of the minimal tissue volume for in vivo measurements, the calibration of a probe constant, the measurement performed on isotropic tissue and the measurement performed on anisotropic tissue. Results showed that the in vitro probe can be used to measure the in-cell dielectric properties of isotropic and anisotropic tissues. When measured with the five in vivo probes, the dielectric properties of isotropic tissue were all measured accurately. For the measurements performed on anisotropic tissue, large errors were observed when the four traditional in vivo probes were used, but only a small error was observed when the new in vivo probe was used. This newly designed five-electrode in vivo probe may indicate the dielectric properties of anisotropic tissue more accurately than these four traditional in vivo probes.

PMID:
23603795
[PubMed - indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Icon for IOP Publishing Ltd.
    Loading ...
    Write to the Help Desk