Prospective clinical study of endoscopic ultrasound-guided choledochoduodenostomy with direct metallic stent placement using a forward-viewing echoendoscope

Endoscopy. 2013;45(5):392-6. doi: 10.1055/s-0032-1326076. Epub 2013 Jan 21.

Abstract

A prospective clinical study was conducted to evaluate the safety, feasibility, and efficacy of endoscopic ultrasound (EUS)-guided choledochoduodenostomy (CDS) with direct metallic stent placement using a prototype forward-viewing echoendoscope. The indication for EUS - CDS in this study was lower biliary obstruction only, and not failed endoscopic biliary drainage, because the aim was to evaluate EUS - CDS for first-line biliary drainage therapy. The technical and functional success rates were 94 % (17 /18) and 94 % (16 /17), respectively. Early complications (focal peritonitis) were encountered in two patients (11 %). No patients developed late complications. EUS - CDS with direct metallic stent placement using a forward-viewing echoendoscope was generally feasible and effective for malignant distal biliary tract obstruction. The forward-viewing echoendoscope was useful, especially for deploying the metallic stent.

Publication types

  • Clinical Trial

MeSH terms

  • Aged
  • Aged, 80 and over
  • Choledochostomy / adverse effects
  • Choledochostomy / instrumentation
  • Choledochostomy / methods*
  • Cholestasis / etiology
  • Cholestasis / surgery*
  • Drainage
  • Endosonography* / adverse effects
  • Feasibility Studies
  • Female
  • Humans
  • Male
  • Middle Aged
  • Neoplasms / complications*
  • Stents
  • Ultrasonography, Interventional* / adverse effects