    Appl Radiat Isot. 2012 Dec;70(12):2737-41. doi: 10.1016/j.apradiso.2012.08.007. Epub 2012 Aug 23.

    An investigation of HPGe gamma efficiency calibration software (ANGLE V.3) for applications in nuclear decommissioning.

    Source

    Department of Physics, University of Surrey, Guildford, UK. s.bell@surrey.ac.uk

    Abstract

    High resolution gamma spectrometry offers a rapid method to characterise waste materials on a decommissioning nuclear site. To meet regulatory requirements, measurements must be traceable to national standards, meaning that the spectrometers must be calibrated for a wide range of materials. Semi-empirical modelling software (such as ANGLE™) offers a convenient method to carry out such calibrations. This paper describes an assessment of the modelling software for use by a small laboratory based on a nuclear site. The results confirmed the need for accurate information on the detection construction if the calibration were to be accurate to within 10%.

    Copyright © 2012 Elsevier Ltd. All rights reserved.

    PMID:
    23041778