    Appl Radiat Isot. 2012 Dec;70(12):2682-5. doi: 10.1016/j.apradiso.2012.08.015. Epub 2012 Sep 15.

    Effects of detector-source distance and detector bias voltage variations on time resolution of general purpose plastic scintillation detectors.

    Source

    Ege University, Faculty of Science, Physics Department, Bornova, Izmir, Turkey.

    Abstract

    Effects of source-detector distance and the detector bias voltage variations on time resolution of a general purpose plastic scintillation detector such as BC400 were investigated. (133)Ba and (207)Bi calibration sources with and without collimator were used in the present work. Optimum source-detector distance and bias voltage values were determined for the best time resolution by using leading edge timing method. Effect of the collimator usage on time resolution was also investigated.

    Copyright © 2012 Elsevier Ltd. All rights reserved.