Radiat Prot Dosimetry. 2013;154(3):331-9. doi: 10.1093/rpd/ncs235. Epub 2012 Sep 5.

A standard for absorbed dose rate to water in a 60Co field using a graphite calorimeter at the National Metrology Institute of Japan.

Author information

  • 1National Metrology Institute of Japan, AIST, Tsukuba, Ibaraki 305-8568, Japan. aist-morishita@aist.go.jp

Abstract

A primary standard for the absorbed dose rate to water in a ⁶⁰Co radiation field has been newly established at the National Metrology Institute of Japan. This primary standard combines the calorimetric measurements using a graphite calorimeter with the ionometric measurements using a thick-walled graphite cavity ionisation chamber. The calorimeter is operated in the constant temperature mode using AC Wheatstone bridges. The absorbed dose rate to water was determined to be 12 mGy s⁻¹ at a point of 1 m from the radiation source and at a water depth of 5 g cm⁻². The uncertainty on the calibration coefficient in terms of the absorbed dose to water of an ionisation chamber using this standard was estimated to be 0.39 % (k=1).