K(3)EDTA Vacuum Tubes Validation for Routine Hematological Testing

ISRN Hematol. 2012:2012:875357. doi: 10.5402/2012/875357. Epub 2012 Jul 24.

Abstract

Background and Objective. Some in vitro diagnostic devices (e.g, blood collection vacuum tubes and syringes for blood analyses) are not validated before the quality laboratory managers decide to start using or to change the brand. Frequently, the laboratory or hospital managers select the vacuum tubes for blood collection based on cost considerations or on relevance of a brand. The aim of this study was to validate two dry K(3)EDTA vacuum tubes of different brands for routine hematological testing. Methods. Blood specimens from 100 volunteers in two different K(3)EDTA vacuum tubes were collected by a single, expert phlebotomist. The routine hematological testing was done on Advia 2120i hematology system. The significance of the differences between samples was assessed by paired Student's t-test after checking for normality. The level of statistical significance was set at P < 0.05. Results and Conclusions. Different brand's tubes evaluated can represent a clinically relevant source of variations only on mean platelet volume (MPV) and platelet distribution width (PDW). Basically, our validation will permit the laboratory or hospital managers to select the brand's vacuum tubes validated according to him/her technical or economical reasons for routine hematological tests.