Factors predicting and having an impact on the need for a permanent pacemaker after CoreValve prosthesis implantation using the new Accutrak delivery catheter system.

Author information

  • 1Department of Cardiology, Virgen de la Victoria Clinical University Hospital, Malaga, Spain. ajmunozgarcia@secardiologia.es

Abstract

OBJECTIVES:

The purpose of this study was to evaluate the need for a permanent pacemaker after transcatheter aortic valve implantation with the CoreValve prosthesis (Medtronic, Inc., Minneapolis, Minnesota) using the new Accutrak delivery system (Medtronic, Inc.).

BACKGROUND:

The need for a permanent pacemaker is a recognized complication after transcatheter aortic valve implantation with the CoreValve prosthesis.

METHODS:

Between April 23, 2008 and May 31, 2011, 195 consecutive patients with symptomatic aortic valve stenosis underwent transcatheter aortic valve implantation using the self-expanding CoreValve prosthesis. In 124 patients, the traditional delivery system was used, and in 71 patients, the Accutrak delivery system was used.

RESULTS:

There were no significant differences in baseline electrocardiographic characteristics between the traditional system and the Accutrak patients: PR interval: 153 ± 46 mm versus 165 ± 30 mm, p = 0.12; left bundle branch block: 22 (20.2%) versus 8 (12.7%), p = 0.21; right bundle branch block: 21 (19.3%) versus 8 (12.7%), p = 0.26. The depth of the prosthesis in the left ventricular outflow tract was greater with the traditional system than with the Accutrak system (9.6 ± 3.2 mm vs. 6.4 ± 3 mm, p < 0.001) and the need for a permanent pacemaker was higher with traditional system than with Accutrak (35.1% vs. 14.3%, p = 0.003). The predictors of the need for a pacemaker were the depth of the prosthesis in the left ventricular outflow tract (hazard ratio [HR]: 1.2, 95% confidence interval [CI]: 1.08 to 1.34, p < 0.001), pre-existing right bundle branch block (HR: 3.5, 95% CI: 1.68 to 7.29, p = 0.001), and use of the traditional system (HR: 27, 95% CI: 2.81 to 257, p = 0.004).

CONCLUSIONS:

The new Accutrak delivery system was associated with less deep prosthesis implantation in the left ventricular outflow tract, which could be related to the lower rate of permanent pacemaker requirement.

Copyright © 2012 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

PMID:
22625192
[PubMed - indexed for MEDLINE]
Free full text
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Elsevier Science
    Loading ...
    Write to the Help Desk