Display Settings:

Format

Send to:

Choose Destination
Med Phys. 2012 May;39(5):2761-4. doi: 10.1118/1.4705366.

Inverse-planned modulated-arc total-body irradiation.

Author information

  • 1Department of Radiation Oncology, University of California, San Francisco, CA, USA.

Abstract

PURPOSE:

To develop a simple and robust method for inverse-planned total-body irradiation (TBI) that is more comfortable and has better dose homogeneity than the conventional forward-planned techniques and that can be delivered in a standard-sized treatment vault.

METHODS:

Modulated-arc TBI (MATBI) utilizes an arc of static open-field beams to irradiate patients as they lay on a stationary couch beneath the gantry, with cerrobend blocks suspended over organs at risk to provide shielding. Prior to treatment, full-body computed tomography (CT) images are acquired of each patient and imported into the PINNACLE(3) planning system, which modulates the monitor units for the open-field beams to optimize the body dose uniformity. The volume of the body within 10% of the prescription dose, V(±10), is used as a metric to evaluate the dose uniformity. For comparison to MATBI, the dose distribution of a conventional forward-planned treatment is also calculated. Quality assurance measurements are acquired before treatment by delivering the plans to a phantom and during treatment with an ionization chamber inside a buildup block, placed between the patient's ankles.

RESULTS:

For MATBI, the achieved values of V(±10) were 75.8%, 90.2%, 84.6%, and 79.8% compared to 60.3%, 77.4%, 65.6%, and 68.5% for the conventional TBI technique, respectively. The pretreatment ion chamber measurements in the phantom had an average error of 1.2%. Those acquired during treatment had larger errors, with most points being within 3% of predictions.

CONCLUSIONS:

MATBI provides better dose uniformity and comfort than the conventional forward-planned TBI techniques. In addition, the technique can be implemented on most linacs, in standard-sized vaults, without the use of a translating couch.

PMID:
22559647
[PubMed - indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Icon for American Institute of Physics
    Loading ...
    Write to the Help Desk