BACKGROUND:

Screening for critical congenital heart defects in newborn babies can aid in early recognition, with the prospect of improved outcome. We assessed the performance of pulse oximetry as a screening method for the detection of critical congenital heart defects in asymptomatic newborn babies.

METHODS:

In this systematic review, we searched Medline (1951-2011), Embase (1974-2011), Cochrane Library (2011), and Scisearch (1974-2011) for relevant citations with no language restriction. We selected studies that assessed the accuracy of pulse oximetry for the detection of critical congenital heart defects in asymptomatic newborn babies. Two reviewers selected studies that met the predefined criteria for population, tests, and outcomes. We calculated sensitivity, specificity, and corresponding 95% CIs for individual studies. A hierarchical receiver operating characteristic curve was fitted to generate summary estimates of sensitivity and specificity with a random effects model.

FINDINGS:

We screened 552 studies and identified 13 eligible studies with data for 229,421 newborn babies. The overall sensitivity of pulse oximetry for detection of critical congenital heart defects was 76·5% (95% CI 67·7-83·5). The specificity was 99·9% (99·7-99·9), with a false-positive rate of 0·14% (0·06-0·33). The false-positive rate for detection of critical congenital heart defects was particularly low when newborn pulse oximetry was done after 24 h from birth than when it was done before 24 h (0·05% [0·02-0·12] vs 0·50 [0·29-0·86]; p=0·0017).

INTERPRETATION:

Pulse oximetry is highly specific for detection of critical congenital heart defects with moderate sensitivity, that meets criteria for universal screening.

FUNDING:

None.

Copyright © 2012 Elsevier Ltd. All rights reserved.

Comment in

PMID:
22554860
[PubMed - indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Icon for Elsevier Science Icon for PubMed Health
    Loading ...
    Write to the Help Desk