Format

Send to:

Choose Destination
See comment in PubMed Commons below
Eur Radiol. 2012 Aug;22(8):1631-40. doi: 10.1007/s00330-012-2424-3. Epub 2012 Apr 19.

Impact of iterative reconstruction on image quality and radiation dose in multidetector CT of large body size adults.

Author information

  • 1Department of Abdominal Imaging and Intervention, Massachusetts General Hospital, Harvard Medical School, 55 Fruit Street, White 270, Boston, MA 02114, USA.

Abstract

OBJECTIVE:

To compare image quality and radiation dose using Adaptive Statistical Iterative Reconstruction (ASiR) and Filtered Back Projection (FBP) in patients weighing ≥ 91 kg.

METHODS:

In this Institution Review Board-approved retrospective study, single-phase contrast-enhanced abdominopelvic CT examinations of 100 adults weighing ≥ 91 kg (mean body weight: 107.6 ± 17.4 kg range: 91-181.9 kg) with (1) ASiR and (2) FBP were reviewed by two readers in a blinded fashion for subjective measures of image quality (using a subjective standardized numerical scale and objective noise) and for radiation exposure. Imaging parameters and radiation dose results of the two techniques were compared within weight and BMI sub-categories.

RESULTS:

All examinations were found to be of adequate quality. Both subjective (mean = 1.4 ± 0.5 vs. 1.6 ± 0.6, P < 0.05) and objective noise (13.0 ± 3.2 vs.19.5 ± 5.7, P < 0.0001) were lower with ASiR. Average radiation dose reduction of 31.5 % was achieved using ASiR (mean CTDIvol. ASiR: 13.5 ± 7.3 mGy; FBP: 19.7 ± 9.0 mGy, P < 0.0001). Other measures of image quality were comparable between the two techniques. Trends for all parameters were similar in patients across weight and BMI sub-categories.

CONCLUSION:

In obese individuals, abdominal CT images reconstructed using ASiR provide diagnostic images with reduced image noise at lower radiation dose.

KEY POINTS:

• CT images in obese adults are noisy, even with high radiation dose. • Newer iterative reconstruction techniques have theoretical advantages in obese patients. • Adaptive statistical iterative reconstruction should reduce image noise and radiation dose. • This has been proven in abdominopelvic CT images of obese patients.