    Cardiovasc Intervent Radiol. 2012 Aug;35(4):942-4. doi: 10.1007/s00270-011-0264-1. Epub 2011 Sep 23.

    Entrapment of the StarClose Vascular Closure System after attempted common femoral artery deployment.

    Source

    Department of Radiology and Biomedical Imaging, University of California, San Francisco, 94143-0628, USA. jeremy.durack@ucsf.edu

    Abstract

    A complication of the StarClose Vascular Closure System (Abbott, Des Plaines, IL) after a transarterial hepatic chemoembolization is described. After attempted clip deployment, the entire device became lodged in the tissues overlying the common femoral artery and could not be removed percutaneously. Successful removal of the device required surgical cutdown for removal and arterial repair. Entrapment of the StarClose vascular closure deployment system is a potentially serious complication that has been reported in the Manufacturer and User Facility Device Experience database, but has not been recognized in the literature.

    PMID:
    21947552
    [PubMed - indexed for MEDLINE]