Measuring the wobble of radiation field centers during gantry rotation and collimator movement on a linear accelerator

Med Phys. 2011 Aug;38(8):4575-8. doi: 10.1118/1.3609098.

Abstract

Purpose: The isocenter accuracy of a linear accelerator is often assessed with star-shot films. This approach is limited in its ability to quantify three dimensional wobble of radiation field centers (RFCs). The authors report a Winston-Lutz based method to measure the 3D wobble of RFCs during gantry rotation, collimator rotation, and collimator field size change.

Methods: A stationary ball-bearing phantom was imaged using multileaf collimator-shaped radiation fields at various gantry angles, collimator angles, and field sizes. The center of the ball-bearing served as a reference point, to which all RFCs were localized using a computer algorithm with sub-pixel accuracy. Then, the gantry rotation isocenter and the collimator rotation axis were derived from the coordinates of these RFCs. Finally, the deviation or wobble of the individual RFC from the derived isocenter or rotation axis was quantified.

Results: The results showed that the RFCs were stable as the field size of the multileaf collimator was varied. The wobble of RFCs depended on the gantry angle and the collimator angle and was reproducible, indicating that the mechanical imperfections of the linac were mostly systematic and quantifiable. It was found that the 3D wobble of RFCs during gantry rotation was reduced after compensating for a constant misalignment of the multileaf collimator.

Conclusions: The 3D wobble of RFCs can be measured with submillimeter precision using the proposed method. This method provides a useful tool for checking and adjusting the radiation isocenter tightness of a linac.

Publication types

  • Evaluation Study

MeSH terms

  • Biophysical Phenomena
  • Humans
  • Movement
  • Particle Accelerators / instrumentation*
  • Particle Accelerators / statistics & numerical data
  • Phantoms, Imaging
  • Radiosurgery / instrumentation*
  • Radiosurgery / statistics & numerical data
  • Radiotherapy, Conformal / instrumentation
  • Radiotherapy, Conformal / statistics & numerical data
  • Rotation