Abstract

BACKGROUND:

During the mid-1990s when our institution was using a press-fit porous-coated cup without supplemental initial fixation for primary THA, the manufacturer transitioned from gamma irradiation to gas plasma for the terminal sterilization of their polyethylene liners.

QUESTIONS/PURPOSES:

At minimum 10-year followup, we asked whether the fixation achieved by solely relying on a press-fit would be durable and how different liner sterilization methods affected radiographic wear, osteolysis, and survivorship.

PATIENTS AND METHODS:

We retrospectively reviewed 373 patients who underwent 398 primary THAs with a press-fit porous-coated cup between March 1995 and December 1996. Mean age at time of surgery was 61.5 ± 13.3 years and mean followup was 10.4 ± 3.7 years. We determined reasons for revision, survivorship, femoral head penetration, osteolysis, and wear-related complications.

RESULTS:

Among 20 revisions involving any component, seven were associated with wear and osteolysis. Kaplan-Meier survivorship, using component revision for any reason as an end point, was 95.7% (95% confidence interval, 93.6%-97.9%) at 10 years. Noncrosslinked liners sterilized with gas plasma demonstrated a mean head penetration rate of 0.20 ± 0.09 mm/year compared with 0.13 ± 0.07 mm/year for liners sterilized with gamma irradiation in air and 0.09 ± 0.04 mm/year for liners sterilized with gamma-irradiation with barrier packaging without oxygen. THAs with increased volumetric wear tended to demonstrate larger osteolytic lesions (r = 0.40) and there tended to be less osteolysis among the liners sterilized with gamma-irradiation with barrier packaging without oxygen. However, there was no difference in survivorship among the sterilization groups and there has been no cup or stem loosening associated with osteolysis.

CONCLUSIONS:

Durable biologic fixation through 10-year followup can be achieved by solely relying on an initial press-fit. Noncrosslinking gas plasma for terminal sterilization of the polyethylene liners was associated with greater head penetration rate than gamma irradiation.

LEVEL OF EVIDENCE:

Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.

PMID:
21904891
[PubMed - indexed for MEDLINE]
PMCID:
PMC3254762
Free PMC Article

Images from this publication.See all images (3)Free text

Fig. 1A–B
Fig. 2
Fig. 3
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Springer Icon for PubMed Central
    Loading ...
    Write to the Help Desk