Monte Carlo simulation of electron beams generated by a 12 MeV dedicated mobile IORT accelerator

Phys Med Biol. 2011 Jul 21;56(14):4579-96. doi: 10.1088/0031-9155/56/14/022. Epub 2011 Jul 1.

Abstract

The aim of this study was to investigate the dosimetric characteristics of the electron beams generated by the light intraoperative accelerator, Liac® (SORDINA, Italy), using Monte Carlo (MC) calculations. Moreover we investigated the possibility of characterizing the Liac® dosimetry with a minimal set of dosimetric data. In fact accelerator commissioning requires measurements of both percentage depth doses (PDDs) and off-axis profiles for all the possible combinations of energy, applicator diameter and bevelled angle. The Liac® geometry and water phantom were simulated in a typical measurement setup, using the MC code EGSnrc/BEAMnrc. A simulated annealing optimization algorithm was used in order to find the optimal non-monoenergetic spectrum of the initial electron beam that minimizes the differences between calculated and measured PDDs. We have concluded that, for each investigated nominal energy beam, only the PDDs of applicators with diameters of 30, 70 and 100 mm and the PDD without an applicator were needed to find the optimal spectra. Finally, the output factors of the entire set of applicator diameters/bevelled angles were calculated. The differences between calculated and experimental output factors were better than 2%, with the exception of the smallest applicator which gave differences between 3% and 4% for all energies. The code turned out to be useful for checking the experimental data from various Liac® beams and will be the basis for developing a tool based on MC simulation to support the medical physicist in the commissioning phase.

MeSH terms

  • Acceleration*
  • Electrons*
  • Intraoperative Period
  • Monte Carlo Method*
  • Radiotherapy / instrumentation*