Format

Send to

Choose Destination
See comment in PubMed Commons below
Comput Math Methods Med. 2011;2011:143480. doi: 10.1155/2011/143480. Epub 2011 Feb 27.

Reducing false alarms of intensive care online-monitoring systems: an evaluation of two signal extraction algorithms.

Author information

  • 1Fakult├Ąt Statistik, Technische Universit├Ąt Dortmund, Dortmund, Germany. borowski@statistik.tu-dortmund.de

Abstract

Online-monitoring systems in intensive care are affected by a high rate of false threshold alarms. These are caused by irrelevant noise and outliers in the measured time series data. The high false alarm rates can be lowered by separating relevant signals from noise and outliers online, in such a way that signal estimations, instead of raw measurements, are compared to the alarm limits. This paper presents a clinical validation study for two recently developed online signal filters. The filters are based on robust repeated median regression in moving windows of varying width. Validation is done offline using a large annotated reference database. The performance criteria are sensitivity and the proportion of false alarms suppressed by the signal filters.

[PubMed - indexed for MEDLINE]
Free PMC Article
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Hindawi Publishing Corporation Icon for PubMed Central
    Loading ...
    Write to the Help Desk