Warning: The NCBI web site requires JavaScript to function. more...
Generate a file for use with external citation management software.
Department of Radiotherapy, Regensburg University Medical Center, Regensburg, Germany. bdobler@web.de
Methods and results for commissioning of the complete VMAT delivery chain are presented for the combination of Nucletron's Oncentra MasterPlan® v3.3 with Elekta's Mosaiq® v1.6 and SynergyS® linac. VMAT specific linac commissioning included determination of the size of the minimal dynamic leaf gap. Dosimetric validation of the complete treatment chain was performed using a 2D-ionization-chamber-array and showed excellent dosimetric results.
Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.
Your browsing activity is empty.
Activity recording is turned off.
Turn recording back on