J Arthroplasty. 2011 Dec;26(8):1527-33. doi: 10.1016/j.arth.2011.01.012. Epub 2011 Mar 11.

Radiographic analysis of a hand-held surgical navigation system for tibial resection in total knee arthroplasty.

Author information

  • 1Adult Reconstruction and Joint Replacement Service, Department of Orthopaedic Surgery, Hospital for Special Surgery, Weill Medical College of Cornell University, New York, New York 10021, USA.

Abstract

Tibial intramedullary or extramedullary alignment guides have not been shown to be highly accurate in performing the tibial resection in total knee arthroplasty (TKA). Since May 2010, a total of 42 knees underwent a TKA using a hand-held, accelerometer-based surgical navigation system for performing the tibial resection (KneeAlign; OrthAlign Inc, Aliso Viejo, Calif). Postoperative standing anteroposterior hip-to-ankle and lateral knee-to-ankle radiographs demonstrated that 97.6% of the tibial components were placed within 90° ± 2° to the mechanical axis in the coronal plane, and 96.2% of the components were placed within 3° ± 2° to the mechanical axis in the sagittal plane. The KneeAlign greatly improves the accuracy of tibial component alignment in TKA.

Copyright © 2011 Elsevier Inc. All rights reserved.

PMID:
21397455
[PubMed - indexed for MEDLINE]