    J Antimicrob Chemother. 2011 Apr;66(4):827-33. Epub 2011 Jan 19.

    Multicentre laboratory validation of the colorimetric redox indicator (CRI) assay for the rapid detection of extensively drug-resistant (XDR) Mycobacterium tuberculosis.

    Source

    Mycobacteriology Unit, Institute of Tropical Medicine, Antwerp, Belgium. amartin@itg.be

    Abstract

    OBJECTIVES:

    To perform a multicentre study to evaluate the performance of the colorimetric redox indicator (CRI) assay and to establish the MICs and critical concentrations of rifampicin, isoniazid, ofloxacin, kanamycin and capreomycin.

    METHODS:

    The study was carried out in two phases. Phase I determined the MIC of each drug. Phase II established critical concentrations for the five drugs tested by the CRI assay compared with the conventional proportion method.

    RESULTS:

    Phase I: a strain was considered resistant by the CRI assay if the MIC was ≥0.5 mg/L for rifampicin, ≥0.25 mg/L for isoniazid, ≥4.0 mg/L for ofloxacin and ≥5.0 mg/L for kanamycin and capreomycin. Sensitivity was 99.1% for isoniazid and 100% for the other drugs and specificity was 97.9% for capreomycin and 100% for the other drugs. Phase II: the critical concentration was 0.5 mg/L for rifampicin, 0.25 mg/L for isoniazid, 2.0 mg/L for ofloxacin and 2.5 mg/L for kanamycin and capreomycin giving an overall accuracy of 98.4%, 96.6%, 96.7%, 98.3% and 90%, respectively.

    CONCLUSIONS:

    Results demonstrate that the CRI assay is an accurate method for the rapid detection of XDR Mycobacterium tuberculosis. The CRI assay is faster than the conventional drug susceptibility testing method using solid medium, has the same turnaround time as the BACTEC MGIT 960 system, but is less expensive, and could be an adequate method for low-income countries.

    PMID:
    21393176