Hypothesis: Navigation can improve accuracy of placement of the glenoid component in reversed shoulder arthroplasty.
Material and methods: A glenoid component of a reversed shoulder prosthesis was implanted in 14 paired scapulohumeral cadaver specimens. Seven procedures with standard instrumentation were compared with 7 procedures using navigation. The intraoperative goal was to place the component centrally in the glenoid in the axial plane and 10° inferiorly tilted in the frontal plane. Glenoid component version and tilt and screw placement were studied using CT scan and macroscopic dissection.
Results: The mean version of the glenoid component in the standard instrumentation group was 8.7° of anteversion, compared with 3.1° of anteversion in the navigated group. The mean tilt of the glenoid component was 0.9° in the standard group and 5.4° of inferior tilt in the navigated group. Using navigation, the range of error for version was 8° (SD 3.3°) compared to 12° (SD 4.1°) in controls. For tilt, the range of error was 8° (SD 3.6°) in navigated specimens and 16° (SD 6.0°) for controls. In the control group, there were no perforations of the central peg, but 1 inferior screw and 4 superior screws were malpositioned. In the navigation group, no central peg perforated, all inferior screws were correctly positioned, and 2 superior screws were malpositioned.
Conclusion: Computer navigation was more accurate and more precise than standard instrumentation in its placement of the glenoid component in reversed shoulder arthroplasty.