    Can J Anaesth. 1990 Apr;37(3):318-21.

    Accuracy of end-tidal PCO2 measurements using a sidestream capnometer in infants and children ventilated with the Sechrist infant ventilator.

    Source

    Department of Anaesthesia, Hospital for Sick Children, University of Toronto, Ontario, Canada.

    Abstract

    To determine the accuracy of end-tidal PCO2 (PETCO2) measurements analyzed with a sidestream capnometer in infants and children whose lungs were ventilated with a Sechrist infant ventilator and an Ayre's t-piece, we compared PETCO2 measurements obtained from the proximal (PETCO2-p) and distal (PETCO2-d) ends of the tracheal tube to arterial PCO2 (PaCO2) in 37 healthy infants and children between 1.3 and 24.5 kg. Both PETCO2-p and PETCO2-d accurately approximated PaCO2, however, the mean (+/- SD) arterial to end-tidal PCO2 difference (delta(a-ET)PCO2) was significantly greater with proximal (1.27 +/- 1.54 mmHg) than with distal sampling (0.64 +/- 1.64 mmHg) (P less than 0.01). In the subgroup of patients who weighted less than 12 kg, the delta(a-ET)PCO2 using proximal gas sampling (1.94 +/- 1.29 mmHg) was also significantly greater than it was using distal sampling (0.74 +/- 1.31 mmHg) (P less than 0.001). We conclude that although statistically different, both proximal and distal estimates of PETCO2 provide acceptable estimates of PaCO2 in healthy infants and children who are ventilated with a Sechrist infant ventilator and an Ayre's t-piece system.

    PMID:
    2108813
    [PubMed - indexed for MEDLINE]

    LinkOut - more resources

    Full Text Sources

    Other Literature Sources

    Molecular Biology Databases

      Supplemental Content

      Icon for Springer

      Save items

      loading

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk