Send to:

Choose Destination
See comment in PubMed Commons below
J Magn Reson Imaging. 2010 Nov;32(5):1255-61. doi: 10.1002/jmri.22356.

Noncontrast SSFP pulmonary vein magnetic resonance angiography: impact of off-resonance and flow.

Author information

  • 1Department of Medicine (Cardiology Division), Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, Massachusetts 02215, USA.



To investigate pulmonary vein (PV) off-resonance and blood flow as causes of signal void artifacts in noncontrast steady-state-free-precession (SSFP) PV magnetic resonance angiography (MRA).


PV blood off-resonance was measured on 11 healthy adult subjects and 10 atrial fibrillation (AF) patients. Noncontrast PV MRA was performed using a 3D slab-selective SSFP sequence at 1.5T on seven healthy subjects with signal profile shifts of 0-125 Hz. The time-resolved blood flow velocity of the PVs was measured on five healthy subjects. The impact of flow was studied on six healthy subjects, on whom SSFP PV MRA was acquired twice with the electrocardiogram (ECG) trigger delay corresponding to low and high flow, respectively.


The PV off-resonances were 97 ± 27 Hz, 65 ± 20 Hz, 74 ± 25 Hz, and 52 ± 17 Hz for right inferior, left inferior, right superior, and left superior PVs, respectively, on healthy subjects, and 74 ± 20 Hz, 38 ± 9 Hz, 51 ± 20 Hz, and 28 ± 11 Hz on AF patients (P<0.01 for all). The off-resonance caused severe signal voids in the PVs. Signal acquired during mid-diastole with high PV flow caused additional signal voids in the left atrium, which was reduced by setting the ECG trigger delay to late-diastole.


PV off-resonance and flow causes signal void artifacts in noncontrast 3D slab-selective SSFP PV MRA.

© 2010 Wiley-Liss, Inc.

[PubMed - indexed for MEDLINE]
Free PMC Article
PubMed Commons home

PubMed Commons

How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Wiley Icon for PubMed Central
    Loading ...
    Write to the Help Desk