    NMR Biomed. 2011 May;24(4):358-65. doi: 10.1002/nbm.1599. Epub 2010 Oct 19.

    Uniform prostate imaging and spectroscopy at 7 T: comparison between a microstrip array and an endorectal coil.

    Source

    University Medical Centre Utrecht, Department of Radiotherapy, the Netherlands. bvandenbergen@hotmail.com

    Abstract

    An endorectal coil and an eight-element microstrip array were compared for prostate imaging at 7 T. An extensive radiofrequency safety assessment was performed with the use of finite difference time domain simulations to determine safe scan parameters. These simulations showed that the endorectal coil can deliver substantially more B(1)(+) to the prostate than can the microstrip array within the specific absorption rate safety guidelines. However, the B(1)(+) field of the endorectal coil is very inhomogeneous, which makes the use of adiabatic pulses compulsory for T(1) - or T(2) -weighted imaging. As a consequence, a full prostate examination is only possible in a feasible amount of time when the microstrip array is used for T(1) - and T(2) -weighted imaging, whereas the endorectal coil is required for spectroscopic imaging. The pulse parameters were optimised within the specific absorption rate guidelines and thereafter used to provide a good illustration of the possibilities of prostate imaging at 7 T.

    Copyright © 2010 John Wiley & Sons, Ltd.

    PMID:
    20960577
    [PubMed - indexed for MEDLINE]

      Supplemental Content

      Icon for John Wiley & Sons, Inc.

      Save items

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk