Display Settings:

Format

Send to:

Choose Destination
See comment in PubMed Commons below

Micro-force sensing in robot assisted membrane peeling for vitreoretinal surgery.

Author information

  • 1ERC for Computer Integrated Surgery, Johns Hopkins University, USA. marcin@jhu.edu

Abstract

Vitreoretinal surgeons use 0.5 mm diameter instruments to manipulate delicate tissue inside the eye while applying imperceptible forces that can cause damage to the retina. We present a system which robotically regulates user-applied forces to the tissue, to minimize the risk of retinal hemorrhage or tear during membrane peeling, a common task in vitreoretinal surgery. Our research platform is based on a cooperatively controlled microsurgery robot. It integrates a custom micro-force sensing surgical pick, which provides conventional surgical function and real time force information. We report the development of a new phantom, which is used to assess robot control, force feedback methods, and our newly implemented auditory sensory substitution to specifically assist membrane peeling. Our findings show that auditory sensory substitution decreased peeling forces in all tests, and that robotic force scaling with audio feedback is the most promising aid in reducing peeling forces and task completion time.

PMID:
20879413
[PubMed - indexed for MEDLINE]
PMCID:
PMC3072686
Free PMC Article

Images from this publication.See all images (5)Free text

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for PubMed Central
    Loading ...
    Write to the Help Desk