Cardiac chamber quantification using magnetic resonance imaging at 7 Tesla--a pilot study

Eur Radiol. 2010 Dec;20(12):2844-52. doi: 10.1007/s00330-010-1888-2. Epub 2010 Jul 17.

Abstract

Objectives: Interest in cardiovascular magnetic resonance (CMR) at 7 T is motivated by the expected increase in spatial and temporal resolution, but the method is technically challenging. We examined the feasibility of cardiac chamber quantification at 7 T.

Methods: A stack of short axes covering the left ventricle was obtained in nine healthy male volunteers. At 1.5 T, steady-state free precession (SSFP) and fast gradient echo (FGRE) cine imaging with 7 mm slice thickness (STH) were used. At 7 T, FGRE with 7 mm and 4 mm STH were applied. End-diastolic volume, end-systolic volume, ejection fraction and mass were calculated.

Results: All 7 T examinations provided excellent blood/myocardium contrast for all slice directions. No significant difference was found regarding ejection fraction and cardiac volumes between SSFP at 1.5 T and FGRE at 7 T, while volumes obtained from FGRE at 1.5 T were underestimated. Cardiac mass derived from FGRE at 1.5 and 7 T was larger than obtained from SSFP at 1.5 T. Agreement of volumes and mass between SSFP at 1.5 T and FGRE improved for FGRE at 7 T when combined with an STH reduction to 4 mm.

Conclusions: This pilot study demonstrates that cardiac chamber quantification at 7 T using FGRE is feasible and agrees closely with SSFP at 1.5 T.

MeSH terms

  • Cardiac-Gated Imaging Techniques / methods*
  • Heart Ventricles / anatomy & histology*
  • Humans
  • Image Enhancement / methods*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Pilot Projects
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Young Adult