    Catheter Cardiovasc Interv. 2010 Jul 1;76(1):60-6.

    An everolimus-eluting stent versus a paclitaxel-eluting stent in small vessel coronary artery disease: a pooled analysis from the SPIRIT II and SPIRIT III trials.

    Source

    Centro Cardiologico Monzino, Istituto di Ricovero e Cura a Carattere Scientifico and Department of Cardiovascular Sciences, University of Milan, Monzino Via Parea 4, Milan, Italy. antonio.bartorelli@ccfm.it

    Abstract

    OBJECTIVES:

    To evaluate the safety and efficacy of the XIENCE V everolimus-eluting stent compared to the TAXUS paclitaxel-eluting stent in small vessels.

    BACKGROUND:

    The XIENCE V everolimus-eluting stent (EES) has been shown to improve angiographic and clinical outcomes after percutaneous myocardial revascularization, but its performance in small coronary arteries has not been investigated.

    METHODS:

    In this pooled analysis, we studied a cohort of 541 patients with small coronary vessels (reference diameter <2.765 mm) by using patient and lesion level data from the SPIRIT II and SPIRIT III studies. TAXUS Express (73% of lesions) and TAXUS Liberté (27% of lesions) paclitaxel-eluting stents (PES) were used as controls in SPIRIT II. In SPIRIT III, Taxus Express(2) PES was the control.

    RESULTS:

    Mean angiographic in-stent and in-segment late loss was significantly less in the EES group compared with the PES group, (0.15 +/- 0.37 mm vs. 0.30 +/- 0.44 mm; P = 0.011 for in-stent; 0.10 +/- 0.38 mm vs. 0.21 +/- 0.34 mm; P = 0.034 for in-segment). EES also resulted in a significant reduction in composite major adverse cardiac events at 1 year (19/366 [5.2%] vs. 17/159 [10.7%]; P = 0.037), due to fewer non-Q-wave myocardial infarctions and target lesion revascularizations. At 1 year, the rate of non-Q-wave myocardial infarction was significantly lower in the EES group compared with that of the PES group (6/366 [1.6%] vs. 8/159 [5.0%]; P = 0.037).

    CONCLUSIONS:

    In patients with small vessel coronary arteries, the XIENCE V EES was superior to the TAXUS PES.

    (c) 2010 Wiley-Liss, Inc.