Anal Biochem. 1991 Jan;192(1):90-5.

Comparison of methods for following alkaline phosphatase catalysis: spectrophotometric versus amperometric detection.

Author information

  • 1Chemistry Department, Oberlin College, Ohio 44074.

Abstract

An amperometric method for alkaline phosphatase is described and compared to the most widely used spectrophotometric method. Catalytic hydrogenation of 4-nitrophenylphosphate (the substrate in the spectrophotometric method) gives 4-aminophenylphosphate (the substrate in the amperometric method). The latter substrate has the formula C6H6NO4PNa2.5H2O and a Mr of 323. The Michaelis constant for 4-aminophenylphosphate in 0.10 M, pH 9.0. Tris buffer is 56 microM, while it is 82 microM for 4-nitrophenyl phosphate. The amperometric method has a detection limit of 7 nM for the product of the enzyme reaction, which is almost 20 times better than the spectrophotometric method. Similarly, with a 15-min reaction at room temperature and in a reaction volume of 1.1 ml, 0.05 microgram/l alkaline phosphatase can be detected by electrochemistry, almost an order of magnitude better than by absorption spectrophotometry. Amperometric detection is ideally suited for small-volume and trace immunoassay.

PMID:
2048739
[PubMed - indexed for MEDLINE]