The advantages of the flow-batch methodology were exploited to implement a simple system with nephelometric detection for the determination of monosodium glutamate (MSG) in food samples. The method is based on the inhibitory effect of the MSG over the crystallization of L-lysine in an isopropanol/acetone mixture. The calibration curve was prepared on-line. The method was linear over the range of 2.8 x 10(-3) to 1.1 x 10(-2)gL(-1) and a detection limit of 9.7 x 10(-5)gL(-1) was achieved. It was successfully applied to determine the MSG concentration in food samples, without a previous treatment. A recovery study was carried out on real samples and the percentages were between 98 and 106%.
(c) 2009 Elsevier B.V. All rights reserved.