Susceptibility-weighted image (SWI) example. Comparison of (a) T2-weighted, (b) SWI filtered phase, (c) processed magnitude, and (d) maximum intensity projection images on patient with traumatic brain injury, acquired on 3 T TRIO Siemens system. SWI has the following acquisition parameters: echo time/repetition time (TR/TE): 29/20 ms, flip angle: 15°, bandwidth: 120 Hz/pixel, 8-channel phased array coil with a parallel imaging factor of two, field of view (FOV): 256 × 256 mm2, slice thickness: 2 mm, acquisition matrix: 512 × 416 × 64, spatial resolution: 0.5 × 0.5 × 2 mm3. T2-weighted image acquired with T2 fast spin echo with TR/TE: 5000/113 ms, FOV: 256 × 256 mm2, slice thickness: 2 mm, acquisition matrix: 320 × 320. Red arrows label multiple possible microhemorrhages invisible on both T1- and T2-weighted images (some not labeled). In this case, SWI data clearly demonstrate multiple possible microhemorrhages in brain.