Format

Send to:

Choose Destination
See comment in PubMed Commons below
Appl Opt. 2009 Dec 10;48(35):6708-15. doi: 10.1364/AO.48.006708.

Optical coherence tomography for quantitative surface topography.

Author information

  • 1Instituto de Optica Daza de Valdés, Consejo Superior de Investigaciones Científicas,C/Serrano 121, 28006 Madrid, Spain. sortiz@io.cfmac.csic.es

Abstract

We address the measurement and correction of fan distortion in optical coherence tomography (OCT). This effect arises from the scanning system configuration and prevents one, in general, from obtaining quantitative topographic data from OCT. Computer simulations allowed us to quantify the effect and evaluate its dependence on the scanning mirror separation and design of the collimating lens, as well as to estimate the optimal axial position of that lens to minimize the fan distortion. We also developed a numerical algorithm based on 3-D ray propagation for the correction of the residual fan distortion. The effect was studied experimentally using a custom developed time-domain OCT in a Michelson configuration provided with a confocal channel, and the accuracy of the fan distortion correction algorithm tested on samples of known dimensions (flat surfaces and spherical lenses). With a proper calibration of the system with use of an onfocal channel, this algorithm makes it possible for time-domain OCT devices to be used as topographers.

PMID:
20011011
[PubMed - indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Optical Society of America
    Loading ...
    Write to the Help Desk