A novel strategy for MALDI-TOF MS analysis of small molecules

J Am Soc Mass Spectrom. 2010 Jan;21(1):154-60. doi: 10.1016/j.jasms.2009.09.024. Epub 2009 Oct 6.

Abstract

Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) does not work efficiently on small molecules (usually with molecular weight below 500 Da) because of the interference of matrix-related peaks in low m/z region. The previous methods developed for this problem focused on reducing the peaks caused by the traditional matrices. Here, we report a novel strategy to analyze small molecules in a high and interference-free mass range by using metal-phthalocyanines (MPcs) as matrices which should be capable of forming matrix-analyte adducts. The mass of the target analyte was calculated by subtracting the mass of MPc from the mass of the MPc-analyte adduct. MPcs were also detectable and could serve as internal standards. Various MPcs with aromatic or aliphatic groups and different metal centers were then synthesized and explored. Aluminum-phthalocyanines (AlPcs), gallium-phthalocyanines (GaPcs), and indium-phthalocyanines (InPcs) were efficient matrices to form MPc-analyte adducts in either the positive or negative ion mode. The detection limits varied from 17 to 75 fmol, depending on analyte types. The mechanism of adducts formation was also proposed. Collectively, our strategy provides a novel and efficient way to analyze small molecules by MALDI-TOF MS.