Format

Send to:

Choose Destination
See comment in PubMed Commons below
Biosens Bioelectron. 2010 Jan 15;25(5):1212-6. doi: 10.1016/j.bios.2009.09.011. Epub 2009 Sep 18.

A biosensor based on photonic crystal surface waves with an independent registration of the liquid refractive index.

Author information

  • 1Institute of Spectroscopy, Russian Academy of Sciences, Troitsk, Moscow Region, Russia. konopsky@isan.troitsk.ru

Abstract

A high-precision optical biosensor technique capable of independently determining the refractive index (RI) of liquids is presented. Photonic crystal surface waves were used to detect surface binding events, while an independent registration of the critical angle was used for accurate determination of the liquid RI. This technique was tested using binding of biotin molecules to a streptavidin monolayer at low and high biotin concentrations. The attained baseline noise is 5x10(-13) m/Hz(1/2) for adlayer thickness changes and 9x10(-8) RIU/Hz(1/2) for RI changes.

Copyright 2009 Elsevier B.V. All rights reserved.

PMID:
19853435
[PubMed - indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Elsevier Science
    Loading ...
    Write to the Help Desk