Representative PSA calibration curve. The calibrator samples were prepared by spiking 0.0, 0.1, 1.0, 5.0, and 10.0 pg/mL of WHO 90:10 PSA standard into human female serum prescreened for PSA (SI Materials and Methods) (Fig. S2). Assuming 100% serum, the final PSA concentrations correspond to a 0.33, 3.3, 16.6, and 33.3 pg/mL calibrator series, accordingly. Samples were run on the automated system with subsequent scanometric detection of the barcode DNA strands released from the 30 nm Au NP probes for PSA target detection and quantification. An example of the gray scale scanometric readout results is shown. The top row of spots in each 2 × 6 well set is the assay response to PSA barcode DNA obtained from the immunoassay, and the bottom set of six spots is the scanometric assay calibrator response. The zero calibrator (no PSA added) is on the far left, with increasing PSA concentrations, as above, moving to the right. The scanometric DNA calibrator sequence is added at the same concentration in each well set.