    Radiat Prot Dosimetry. 2009;136(3):203-8.

    Patient dose assessment due to fluoroscopic exposure for some selected fluoroscopic procedures in Ghana.

    Gyekye PK, Schandorf C, Boadu M, Yeboah J, Amoako JK.

    Ghana Atomic Energy Commission, Radiation Protection Institute, PO Box LG80, Legon, Accra. pgyekye@yahoo.com

    Organ and effective doses to 90 patients undergoing some selected fluoroscopic examinations at the Korle-Bu Teaching Hospital were estimated using the Monte Carlo-based program (PCXMC version 1.5). Radiation dose was estimated from free-in-air measurements. The mean effective doses were found to be 0.29 +/- 0.07, 0.84 +/- 0.13, 3.15 +/- 0.44, 6.24 +/- 0.70 and 0.38 +/- 0.05 mSv for urethrogram, barium swallow, barium meal, barium enema and myelogram examinations, respectively. The dose area product was estimated to be 3.55 +/- 0.95, 16.44 +/- 2.60, 50.81 +/- 7.04, 99.69 +/- 10.85 and 9.32 +/- 0.99 Gy cm(2) for urethrogram, barium swallow, barium meal, barium enema and myelogram examinations, respectively. Optimisation of procedures is required for barium enema and barium meal examinations.

    PMID: 19770212 [PubMed - in process]