Measurement performance of reagent manufacturers by Centers for Disease Control and Prevention/Cholesterol Reference Method Laboratory Network lipid standardization specified for metabolic syndrome-focused health checkups program in Japan

J Atheroscler Thromb. 2009;16(6):756-63. doi: 10.5551/jat.1503. Epub 2009 Sep 18.

Abstract

Aim: This study was designed to clarify the current measurement performance of 7 reagent manufacturers for high-density lipoprotein cholesterol (HDLC), low-density lipoprotein cholesterol (LDLC) and triglycerides (TG) specified for the metabolic syndrome (MetS)-focused health checkups program in Japan.

Methods: Twenty HDLC, 21 LDLC and 9 TG analytical reagent/instrument/calibrator systems (system), and combinations of reagent lots, instrument models and calibrator lots, underwent Centers for Disease Control and Prevention (CDC)/Cholesterol Reference Method Laboratory Network (CRMLN) lipid standardization. Eighty and 100% systems were requested to achieve an accuracy of within +/-1% and +/-2% of the reference value, so that a clinical laboratory can meet the CDC criteria.

Results: The CDC performance criteria of HDLC, LDLC and TG require an accuracy of within +/-5%, +/-4% and +/-5%, respectively. For HDLC, all 20 systems met the criteria. Fourteen (70.0%) and 18 (90.0%) systems were within +/-1% and +/-2%, respectively. For LDLC, 14 (66.7%) of 21 systems met the criteria, but 7 (33.3%) failed. Five (23.8%) and 17 (81.0%) systems were within +/-1% and +/-2%, respectively. For TG, 8 of 9 systems met the criteria. Two (22.2%) and 4 (44.4%) systems were within +/-1% and +/-2%, respectively. The minimum and maximum differences of a specified sample among manufacturers were 1.6 and 11.0 mg/dL for HDLC, 7.8 and 33.0 mg/dL for LDLC, and 2.8 and 27.4 mg/dL for TG, respectively.

Conclusion: Homogeneous HDLC methods are acceptable for MetS, but further accuracy improvement of homogeneous LDLC and TG methods will be needed because of their poor performance.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Chemical Analysis / standards
  • Calibration / standards
  • Centers for Disease Control and Prevention, U.S. / standards
  • Cholesterol / blood
  • Cholesterol / standards*
  • Cholesterol, HDL / blood
  • Cholesterol, HDL / standards*
  • Humans
  • Japan
  • Laboratories / standards
  • Lipids / chemistry
  • Metabolic Syndrome / blood
  • Metabolic Syndrome / diagnosis*
  • Quality Control
  • Reagent Kits, Diagnostic / standards*
  • Reference Values
  • Reproducibility of Results
  • United States

Substances

  • Cholesterol, HDL
  • Lipids
  • Reagent Kits, Diagnostic
  • Cholesterol