Diagnostic performance of a novel device for real-time margin assessment in lumpectomy specimens

J Surg Res. 2010 May 15;160(2):277-81. doi: 10.1016/j.jss.2009.02.025. Epub 2009 Mar 31.

Abstract

Background: Margin status in breast lumpectomy procedures is a prognostic factor for local recurrence and the need to obtain clear margins is often a cause for repeated surgical procedures. A recently developed device for real-time intraoperative margin assessment (MarginProbe; Dune Medical Devices, Caesarea, Israel), was clinically tested. The work presented here looks at the diagnostic performance of the device.

Methods: The device was applied to freshly excised lumpectomy and mastectomy specimens at specific tissue measurement sites. These measurement sites were accurately marked, cut out, and sent for histopathologic analysis. Device readings (positive or negative) were compared with histology findings (namely malignant, containing any microscopically detected tumor, or nonmalignant) on a per measurement site basis. The sensitivity and specificity of the device was computed for the full dataset and for additional relevant subgroups.

Results: A total of 869 tissue measurement sites were obtained from 76 patients, 753 were analyzed, of which 165 were cancerous and 588 were nonmalignant. Device performance on relatively homogeneous sites was: sensitivity 1.00 (95% CI: 0.85-1), specificity 0.87 (95% CI: 0.83-0.90). Performance for the full dataset was: sensitivity 0.70 (95% CI: 0.63-0.77), specificity 0.70 (95% CI: 0.67-0.74). Device sensitivity was estimated to change from 56% to 97% as the cancer feature size increased from 0.7 mm to 6.6 mm. Detection rate of samples containing pure DCIS clusters was not different from rates of samples containing IDC.

Conclusions: The device has high sensitivity and specificity in distinguishing between normal and cancer tissue even down to small cancer features.

Publication types

  • Clinical Trial
  • Multicenter Study
  • Validation Study

MeSH terms

  • Adult
  • Biopsy
  • Breast / pathology
  • Breast / surgery
  • Breast Neoplasms / pathology*
  • Breast Neoplasms / surgery*
  • Female
  • Humans
  • Intraoperative Care / instrumentation
  • Intraoperative Care / standards
  • Mastectomy, Segmental*
  • Pathology, Clinical / instrumentation*
  • Pathology, Clinical / standards
  • ROC Curve
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectroscopy, Near-Infrared / instrumentation*
  • Spectroscopy, Near-Infrared / standards