J Biomed Opt. 2009 May-Jun;14(3):034017. doi: 10.1117/1.3147402.

New cross-talk measure of near-infrared spectroscopy and its application to wavelength combination optimization.

Author information

  • 1National Institute of Advanced Industrial Science and Technology, Neuroscience Research Institute, 1-1-1 Umezono Central 2, Tsukuba-shi, Ibaraki 305-8568, Japan. s.umeyama@aist.go.jp

Abstract

In near-infrared spectroscopy, concentration changes in oxygenated and deoxygenated hemoglobin are calculated from the changes in the attenuation of the measurement light. This is done by solving a linear equation based on the modified Lambert-Beer law. To solve this equation, we need to know the partial optical pathlengths in the activated region in the brain. Because they are difficult to know, a wavelength-independent constant or a wavelength-dependent total optical pathlength has been substituted for these values in actual measurements. This kind of substitution inevitably produces errors, called cross-talk, when calculating concentration changes. In this paper, we propose a new cross-talk measure for dual and triple wavelength measurements, and analyze it over various wavelength combinations. The results indicate that constant substitution is not inferior to total path-length substitution in dual wavelength measurements, and that total path-length substitution is very effective for triple wavelength measurements.

PMID:
19566310
[PubMed - indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Icon for Silverchair Information Systems
    Loading ...
    Write to the Help Desk